MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex patient lift. Regular
maintenance of your patient lift is necessary to ensure continuing proper and safe opera-
tion. Read and observe the following recommended maintenance schedule:
Item Inspect
initially
Inspect/adjust
daily monthly every 6
months
Base and Casters
Ensurethatbaseislevel(allfourwheelstouchtheoor) X X X
Ensurethatcastermountingboltsandaxleboltsaretight X X
Ensurethatcastersswivelandrollfreely X X
Applyalightgreasetocasterballbearings X
Base Adjustments
Ensurethatbaseopensandclosessmoothly X X X
Ensurethatbaselockssecurelyinthreepositions X X
Ensurethatlockingknobisinplaceandoperatingproperly X X
Inspectallbasehardwareforwear X
Mast
Ensurethatmastisfullyengagedandlockedintomastsleeve X X
Inspectmastforbendsordeections X
Inspectallmasthardwareforwear X X
Boom
Inspectboomforbendsordeections X X
Ensurethatboomiscenteredbetweenthebaselegs X X X
Inspectallboomhardwareforwear X
Spreader Bar
Inspecthangerpostforwearordamage X
Inspectslinghookforwearordeections X X
Inspectallspreaderbarhardwareforwear X X
Slings
Inspectslingsandstrapsforwearordamage X
Inspectmetal/plasticpartsforwearordamage X
Lubrication with light oil
Lubricateallpivotpoints X
Lubricatetopandbottomofactuatormounting X
Cleaning
Cleanwithmildnon-abrasivedetergent X
Pump
Ensurethatpumpmountinghardwareistight X X
Inspectpumpmountinghardwareforwear X